IND780 User’s Guide
PDX Scale Data (POWERCELL PDX bases only)
The PDX Scale Data page (Figure 3-45) shows individual load cell information, for
a PDX type scale. This page provides access to several views:
• Adjusted Counts • Tilt and Energy (MT Service Only)
• Gross Weight • Input Voltage (MT Service only)
• Communication Errors • Gas Concentration (MT Service Only)
• Temperature (MT Service only)
As indicated in the list above, some views are accessible only if the terminal’s MT
Service Security has been unlocked.
In the data pages each cell node is represented by a circle, with different colors to
indicate if certain fault conditions exists. A circle with a red border (as in cells 2
and 7 in the example below) indicates that a fault condition has been detected, but
in a different view. A completely red circle indicates that a fault is present in the
current view.
Indicated fault conditions include cell communication errors, cell zero drift, cell
overloads and enclosure breach. Click on each individual cell to go to a specific
detailed PDX Load Cell page (Figure 3-47).
Figure 3-45: PDX Scale Data Page
PDX Terminal (POWERCELL PDX bases only)
The PDX Terminal page (Figure 3-46) shows the minimum and maximum
voltages and current detected on the IND780 PDX option board. This page can be
used to determine if the terminal’s PDX option board is providing the proper
electrical parameters to support the CAN Bus communication and to power the
cells. It also helps to indicate if there are issues with poor wiring, a short condition
or an overloaded power supply. Most of the information on this page is only
accessible if the terminal’s MT Service Security has been unlocked.