IND780 User’s Guide
2. Access Setup > Maintenance > Run > Backup to USB. The screen shown in
Figure 3-6 will appear.
Figure 3-6: Backup to USB
3. By default, the Dataset Name is BK1; press ENTER to select the field and enter
another name, if required.
4. Press the START softkey
. The IND780 will check for the presence of a valid
USB memory device. On finding one, it checks for available space. If a device
is not found, or a device is present that has insufficient space for the operation,
a prompt (“USB Memory Error”) informs the operator and the backup does not
proceed. If space is available, the IND780 creates a folder titled Mettler
Toledo/Backup/IND780/ <dataset name > (where “dataset name” is the name
entered in step 3), then begins the transfer of data. An on-screen message
(“Working”) indicates that the procedure is in progress.
5. Once the terminal indicates that the backup is complete, the USB device can
be unplugged and the EXIT softkey
pressed to return to the menu tree.
Restore
System Restore cannot be performed if the terminal is locked.
An Administrator level login must be in effect to carry out this procedure.
A system restore operation does not copy log files from the dataset, and the
terminal’s existing log files are not over-written on restore.
To restore files to an IND780:
1. Connect a USB (flash) memory device to the IND780’s USB port, either directly
or via a USB extension cord.