IND780 User’s Guide
WARNING!
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THE TERMINAL.
EXERCISE CARE WHEN MAKING CHECKS, TESTS, AND ADJUSTMENTS
THAT MUST BE MADE WITH POWER ON. FAILING TO OBSERVE THESE
PRECAUTIONS CAN RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
Troubleshooting
Troubleshooting activities for the IND780 include:
• Board LED Interpretation • Internal diagnostic testing
• AC power test • MT Service View
• Power supply voltage test • External diagnostics
• Battery test • Watchdog monitoring
• RS-232 serial output voltage test • Master reset
Troubleshooting Overview
Each of the procedures listed in Table 3-1 is described in detail in the following
sections. If the result of any of these procedures indicates a problem, refer to this
table to isolate the possible cause/s and to identify the appropriate response.
Table 3-1: Troubleshooting
Diagnostic
Procedure
Expected Outcome Response/s to Incorrect Outcome
AC Power Test
Power source within
range -5% to +10% of
rated VAC (100-
240 VAC, depending on
terminal configuration)
If display operates incorrectly, or there are
intermittent problems with Terminal function:
• Check whether power source is within
correct range
• If power source is not OK, find a reliable
power source
• If problem persists, perform Power
Supply Voltage Test
Power Supply
Voltage Test
12 VDC: In range 11.7-
12.3 VDC
5 VDC: In range 4.9-
5.1 VDC
If voltages are absent or outside specified
ranges, replace the Power Supply
Battery Test Not less than 2.5 VDC
If voltage is below 2.5 VDC, replace battery
BRAM Test BRAM PASS
If BRAM FAIL displays, replace Main PCB
3-10