Figure 2-16 shows the Home screen after the Unit Switching softkey has been
pressed, displayed the gross weight in pounds.
Figure 2-16: Home Screen After the Unit Switching Softkey has Been Pressed
IDNet Class II
If an IDNet scale base is set up as a Class II W&M approved scale with the
appropriate capacity and increment size, the IND780 will show a weight display
with the last digit in a smaller size (Figure 2-17).
Figure 2-17: Class II-Approved IDNet Base Weight Display
A demand print of weight data in this case will include parentheses around the
last digit of each weight field, to indicate the Class II approval condition.
Example:
Gross: 2.767(8) kg
Tare: 1.719(3) kg T
Net: 1.048(5) kg
Expand By 10
The EXPAND BY 10 softkey is used to increase the selected weight display
resolution by one additional digit. For example, a weight display of 40.96 could