Mettler Toledo MT Guide de l'utilisateur Page 44

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 129
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 43
IND780 User’s Guide
Figure 2-12: Summed Scales Over Capacity (left) and Under Range (right)
If one scale included in the Sum Scale is over-capacity and another is under range,
the respective scales will indicate their conditions accordingly, and the Sum Scale
display will show “----------“, indicating an invalid weight value. This condition is
shown in Figure 2-13.
Figure 2-13: Summed Scales Simultaneously Over Capacity and Under Range
Zero
The Zero function is used to set or reset the initial zero reference point of the
IND780. There are three types of zero setting modes:
Automatic Zero Maintenance (AZM)
Power-Up
Pushbutton
Automatic Zero Maintenance
Automatic Zero Maintenance (AZM) enables the IND780 to compensate for the
buildup of small amounts of weight and track itself back to the center of zero.
Within the AZM operating range (programmable from 0.0 to 9.9 divisions), when
the terminal is in a no motion condition it makes small adjustments to the current
zero reading to drive the weight reading toward the true center-of-zero. When the
weight is outside of the programmed AZM range, this feature is not functional.
2-19
Vue de la page 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 128 129

Commentaires sur ces manuels

Pas de commentaire