IND780 User’s Guide
user to track changes and make data comparisons to a known working
condition of the cell.
Figure 3-36: Load Cell Information Screen
The recorded load cell parameters include:
Data Field Description
Cell S/N
Represents the unique factory serial number embedded in the cell.
Install Date
Indicates the date the load cell was last addressed.
Temperature
Temperature sensor reading in the cell.
LC Supply Voltage
Input supply voltage measured in the cell in volts.
Displacement
Cell angular displacement in degrees.
Energy
Cell tilt energy in (milliradians/sec)
2
.
CAN-High Dominant X
Cell CAN-High voltage recorded for Dominant mode in volts.
1
CAN-Low Dominant X
Cell CAN-Low voltage recorded for Dominant mode in volts.
1
CAN-High Recessive
Cell CAN-High voltage recorded for Recessive mode in volts.
1
CAN-Low Recessive
Cell CAN-Low voltage recorded for Recessive mode in volts.
1
Gas Concentration
Level (%) of inert gas concentration enclosed in the cell as a
measure of the hermetic sealing. Small losses of gas levels are
expected over the years. A sudden rapid loss of gas levels is an
apparent indication of a cell enclosure breach condition.
Note
1. These values are stored from the last time the Load Cell COM Voltage Screen was
accessed.
PDX Terminal — From the MT Service View screen, choose Terminal in the
Device selection box and press the VIEW softkey
to display the PDX
Terminal screen. This screen provides a view of the minimum and maximum
voltages and current detected on the IND780 PDX option board. These values
are updated dynamically while the screen is viewed.
3-29