IND780 User’s Guide
Figure 3-21: Calibration Values Screen, Zero in Focus
Shift Values—Used for POWERCELL and PDX scales only. The following
screen shows values used to adjust gain for each load cell in a multiple load
cell system. The values compensate for difference in response between cells.
The load cell numbers represent the node cell addresses. Where more than 12
load cells are present this screen displays a scroll box to reveal further rows.
The diagnostic purpose of the Load Cell Output and Shift Values screens is to
permit a technician to confirm valid calibration values were obtained when the
calibration routine (Setup > Scale > Scale
n
> Calibration) was carried out.
Figure 3-22: Scale Shift Values Screen
Diagnose Cell—Used for POWERCELL scales only (but not POWERCELL PDX).
The purpose of the Diagnose Cell screens is to allow the technician to verify a
load cell’s node address and raw counts. If the cell that is being diagnosed
has a known address, select that address from the Load Cell Address selection
box. If the cell address is unknown, then select Connected Cell. If the cell is
faulty or it has an invalid address, the terminal will fail to communicate with
the cell. If the cell responds correctly, the node address and the cell counts will
be displayed on the screen (Figure 3-23).