Chapter 5: General Installation Guidelines
Structural Support Guidelines
(12/99)
• Figure 5-8c: Support structure is out of level, applying side forces to the load cell.
A tank scale’s support structure should deflect as little as possible, and any
deflection should be uniform at all support points (see Figure 5-9). Excessive
deflection can cause inlet and outlet piping to bind, creating linearity errors. When
deflection is not uniform, it can cause repeatability errors and zero return errors
due to creep.
1/4” max.
Figure 5-9: Weigh Module Base Support Structure Deflection
In some cases, a tank’s legs will deflect under the weight of the tank (see Figure
5-10a). If the deflection is great enough to affect weight readings, you should
brace the legs to keep them rigid (see Figure 5-10b).
Figure 5-10a: Deflection of Tank Legs Figure 5-10b: Tank Legs Braced
Minimize Deflection
Commentaires sur ces manuels