Mettler Toledo LJ16 Spécifications Page 134

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 198
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 133
METTLER TOLEDO Weigh Module Systems Handbook
(12/99)14-14
Appendix 7: Load Cell
Deflection
Flexmount/Centerlign Flexmount/Centerlign Ultramount Tension Mount Tension Mount
Load Cell
Capacity
(lb)
Deflection
at Rated
Capacity
(inches)
Load Cell
Capacity
(kg)
Deflection
at Rated
Capacity
(mm)
Load Cell
Capacity
(kg)
Deflection
at Rated
Capacity
(mm)
Load Cell
Capacity
(lb)
Deflection
at Rated
Capacity
(inches)
Load Cell
Capacity
(kg)
Deflection
at Rated
Capacity
(mm)
250 0.010 220 0.254 5 <0.4 50 0.006 50 0.152
500 0.010 550 0.254 10 <0.4 100 0.006 100 0.152
1,250 0.010 1,100 0.254 20 <0.4 200 0.006 200 0.152
2,500 0.010 2,200 0.254 50 <0.4 300 0.007 500 0.152
5,000 0.012 4,400 0.254 100 <0.4 500 0.007 1,000 0.254
10,000 0.012 9,072 0.910 1,000 0.006 2,000 0.381
20,000 0.036 13,608 0.910 2,000 0.010 5,000 0.381
30,000 0.036 20,412 1.650 3,000 0.012
45,000 0.065 5,000 0.015
50,000* 0.020 10,000 0.015
75,000* 0.020 ——————
100,000* 0.020 ——————
150,000* 0.020 ——————
200,000* 0.020 ——————
*50,000-lb to 200,000-lb capacities are used only for Flexmount HD weigh modules.
Table 14-6: Load Cell Deflection