Chapter 5: General Installation Guidelines
Tank and Vessel Design
(12/99)
5-5
air is vented, pressure will build up inside the tank. The increased pressure will
produce a weighing error until a pressure balance can be restored. A similar
condition occurs when a material is discharged rapidly from a tank, creating a
partial vacuum inside the tank. To prevent pressure imbalances, make sure that
the tank is adequately vented. That will allow you to weigh the contents accurately
as soon as the tank is filled or emptied, instead of having to wait for a pressure
balance to be restored. Vents should be vertical and provided with clean-out
doors and fume stops or dust collectors.
Provisions for Test Weights
If you are going to use test weights to calibrate a tank scale, you will need some
way to hang the test weights from the tank. In most cases, this can be done with
some type of mounting lugs spaced evenly around the tank. Figure 5-7 shows a
mounting lug with a test weight hanging from it. Use a hoist for raising/lowering
the weight.
Figure 5-7: Mounting Lugs for Test Weight
Mounting Lug
Test Weight
Hoist/Jack
Chain or
Nylon Strap