Section 14: Technical data and accessories
105
14 Technical data and accessories
In this Section you will find the most important technical data for your balance. Accessories from the METTLER TOLEDO range increase
the functionality of your balance and open up additional areas of application. In this Section you will find a list of the options currently
available.
14.1 General data
Power supply
Power supply connector 11100750, KA-52A
Primary: 100-240V, -15%/+10%, 50/60Hz, 0.7A
with AC/DC adapter: Secondary: 12VDC +/-5%, 2.08A (with electronic overload protection)
or
11106930, PSU30A-3
Primary: 100-240V, -15%/+10%, 50/60Hz, 0.8A
with AC/DC adapter: Secondary: 12VDC +/-5%, 2.25A (with electronic overload protection)
Power supply cable: 3-core, with country-specific plug
Power supply to the balance: 12 VDC +/-5%, 2.08A, maximum ripple 120 mVpp
Use only with a tested AC adapter with SELV output current.
Ensure correct polarity
Protection and standards
Overvoltage category: Class II
Degree of pollution: 2
Protection: Protected against dust and water
Standards for safety and EMC: See Declaration of Conformity (separate brochure 11780294)
Range of application: For use only in closed interior rooms
Environmental conditions
Height above mean sea level: Up to 4000 m
Ambient temperature: 5-40 °C
Relative air humidity: Max. 80% at 31°C, linearly decreasing to 50% at 40 °C, noncondensing
Warm-up time: At least 120 minutes after connecting the balance to the power supply; when
switched on from standby-mode, the balance is ready for operation immediately.
Materials
Housing: Die-cast aluminum, lacquered
Terminal: Die-cast zinc, lacquered, and plastic
Weighing pan: Chrome steel
Standard equipment
Delivered with balance: AC adapter with country-specific power cable
RS232C interface
Protective cover for the terminal
Feedthrough for below-the-balance weighing, Antitheft device (AX models only)
Cleaning brush, Cleaning tweezers (MX/UMX models only)
Weighing tweezers (AX26, AX26DR, AX205, AX205DR, MX5, UMX2, and UMX5 only)
Documentation: Operating instructions and booklet “Weighing the right way with METTLER TOLEDO”
Production certificate