7 Malfunctions
7.1 Interpreting the error messages
Malfunction/Message Possible cause Rectification
Printout unreadable Ribbon worn/entangled Change/tension ribbon
Service life of printing unit at an end Replace printing unit
Green pilot lamp does not light up No power, fuse defective Switch on power switch, have fuse replaced by
METTLER TOLEDO
Green pilot lamp flashes slowly Balance waiting for stability/
cable not attached
Green
pilot lamp flashes rapidly Transmission error
Connection cable not attached or Check handshake, initiate print command
older balance (unidir. interface) with PRINT key of balance
Interface parameters of LC-P45 Set interface parameters
and balance do not match
Calibration has been aborted Repeat calibration
------- ABORT --------
---- Interruption ----
- Transmission error -