20
– When necessary enter sample number start value (page 12)
Enter the factor z (ISO 8655) to calculate the volume from the mass (page 12). This factor is es-
sentially
a function of the water temperature and atmospheric pressure
Enter the rounding off convention for the calculated result (page 12)
Dosing the liquid and inclusion of the first volume in the statistics. The dosed volume is calculated
automatically from the measured weight by multiplying it with the factor z. The weight and the cal-
culated
volume in µl, (V = m x z) are printed out
Continue the sample series
If desired, print out intermediate statistics (the sample counter is not reset) (page 16)
Continue the sample series
Conclusion of the sample series with automatic printout of the overall statistics (the sample counter
is reset, the statistics memory cleared) (page 16)
1
F
2
Result
Clear
:
:
:
:
VoluMate
L I Q UI S Y S T EM S
®
...
7
Def
6
21 3
...
7
Def
...
7
Def
2
Result