12
METTLER TOLEDO US and Canada
Commonly used techniques
Electrolytic conductivity is a measure
of the total ionic content of a solution.
Two main methodologies to measure
conductivity are:
2-electrode sensors are for measure-
ments in high purity water and
relatively low conductivity ranges
4-electrode sensors are for mid to
high ranges. They are more resistant
to fouling than two-electrode designs
METTLER TOLEDO offers both method-
ologies.
2-electrode sensor design (Fig.1)
An AC voltage is applied across the 2
electrodes, and the resistance between
them is measured. The built-in
temperature sensor provides fast accu-
rate measurement. The cell geometry
and the high solution resistance allow
for very accurate and precise conduc-
tivity determination.
Sensors are used for water condition-
ing and purification stages where they
are capable of detecting minute levels
of impurities in ultrapure water.
4-electrode sensor design (Fig.2)
An AC voltage is applied across the
2 outside electrodes. The principle is
to measure the voltage drop across the
2 inner electrodes. Therefore, polariza-
tion errors are eliminated. Since this
technique measures potential drop,
the measurement remains accurate. It
permits easier in-line cleaning and it
can be installed in smaller piping than
inductive sensors.
These sensors are used for concentra-
tion measurement of acids, alkalis,
and salt process streams.
UniCond
®
Conductivity/Restivity
Sensors with ISM
®
(Fig.3)
The UniCond conductivity/resistivity
sensor advancement integrates the
measuring circuit and the physical
sensor into a single unit. UniCond
conductivity/resistivity sensors pro-
vide exceptionally wide measurement
ranges due to their advanced built-in
measuring circuit. The on-board meas-
uring circuit eliminates interference
from lead wire resistance and capaci-
tance. Only digital signals go back to
the transmitter. The UniCond design
mitigates the effects of polarization
enabling the upper range of the
conductivity sensor to be greatly
expanded. UniCond sensors provide
the ability to accurately measure from
ultrapure water to brackish water (up
to 50,000 µS / cm) with a single inte-
grated sensor, greatly simplifying water
treatment instrumentation.
Conductivity / Resistivity Systems
When Optimal Performance Is Essential
Inner
electrode
rod
Outer electrode cylinder
Fig. 1: Concentric 2-electrode design
Measuring
electrodes
AC voltage
measurement
Drive electrodes
Four electrode
sensor
AC voltage
source
Fig. 2: 4-electrode sensor design
Fig. 3: UniCond
®
Conductivity / Resistivity
Introduction
Control Your Process Regardless of Your Application
Electrolytic conductivity is a widely used analytical parameter for water purity analysis,
monitoring of reverse osmosis, cleaning procedures, control of chemical processes, and in
industrial wastewater.
Vue de la page 0
1 2 3 4 5 6 ... 13 14

Résumé du contenu

Page 1 - Conductivity / Resistivity

12METTLER TOLEDO US and CanadaCommonly used techniquesElectrolytic conductivity is a measure of the total ionic content of a solution. Two main method

Page 2

21METTLER TOLEDO US and Canada 1.00 mm 25.4"27.9 mm1.10"101.6 mm4"1" NPT76.2 mm3.00"  44.5 mm1.75"¼" NPT¼&qu

Page 3

22METTLER TOLEDO US and CanadaPharma Waters VerifiersConductivity / Resistivity SensorsAccurate and ReliablePharma Water VerifiersUniCond® ISM Verifi

Page 4

23METTLER TOLEDO US and CanadaPharma WaterOrdering InformationDescription Order NumberPharma Waters UniCond® ISM Verifier for Thornton M800 and M30

Page 5

24METTLER TOLEDO US and Canadawww.mt.com / ThorntonWater QualityPure LightSimple Water Quality Indication656 Pure Light™Water Quality Indicator with T

Page 6

25METTLER TOLEDO US and CanadaSpecificationsAccuracy ± 10% at 25 °C, typicalThreshold settings Restivity 50 k, 200 k, 500 k, 1 M &

Page 7

13METTLER TOLEDO US and Canada58 031 409243E233240-201Continuous conductivity monitoring according to USP <645>USP guideline <645> sets a

Page 8

14METTLER TOLEDO US and CanadaFeatures Overview– Plug and Measure functionality– Integral high-performance measuring circuit– Robust digital output s

Page 9

15METTLER TOLEDO US and Canada3/4" NPTX[ 15.7 mm0.62"[ 12.7 mm0.50"[ 15.7 mm0.62"[ 104 mm / 4.1"[ 109 mm / 4.3"[ 7

Page 10 - US and Canada

16METTLER TOLEDO US and Canadawww.mt.com / Thornton-CondConductivity / ResistivityMETTLER TOLEDO Thornton provides a full complement of conductivity /

Page 11 - Pharma Waters Verifiers

17METTLER TOLEDO US and CanadaConductivity Standard SolutionsProvided for sensor verification and recalibration, conductivity standards are pro-duced,

Page 12

18METTLER TOLEDO US and Canadawww.mt.com / Thornton-CondConductivity / ResistivityConductivity / Resistivity SensorsAccurate and ReliableConductivity

Page 13 - Water Quality

19METTLER TOLEDO US and CanadaMicroelectr.PowerPharma WaterWastewaterOrdering InformationElectrode Maximum Pressure Process Connection Insertion