6. What if ...?
Possible cause
Spacer bolt too short
Draft shield not mounted correctly
Compensation weight not installed in draft
shield
Spacer bolt too long
Draft shield not mounted properly
•Weighing range exceeded
Spacer bolt too long
Power outage
Instrument not switched on
No power supply voltage
Power cable not plugged in
Voltage selector not set properly
Microfuse blown
Balance switched off
Connection cable from balance to dryer
loose or not plugged in
The configuration has been changed
Rectification
Lengthen spacer bolt
•Mount draft shield correctly
Install compensation weight
Shorten spacer bolt
•Mount draft shield correctly
Remove sample
shorten
Repeat drying operation. The parameter
entries are not lost through a power out-
age.
Switch on dryer
Check power supply
Plug in power cable
Set voltage selector correctly
Check functioning of the microfuses and
replace fuses if need be
Switch on balance
Plug in connection cable properly
Reenter operating parameters
Error/Display
After switching on, only the bottom segments
of the display light up
Unstable value in the display
After addition of the sample, only the top
segments of the display light up
Display “OFF”
No LED light up on the dryer
Only the LED START flashes
After switching on, the configuration is wrong